TITLE

Radiographic Distortion Artifact of Circular External Fixators

AUTHOR(S)
Secrest, Scott; Nagy, Jesse; Kneller, Stephen
PUB. DATE
May 2015
SOURCE
Journal of the American Animal Hospital Association;May/Jun2015, Vol. 51 Issue 3, p143
SOURCE TYPE
Academic Journal
DOC. TYPE
Article
ABSTRACT
Circular external fixators are commonly used to surgically treat a variety of orthopedic conditions. However, distortion artifact may mislead the radiographic interpreter as to the true anatomic location of the transfixing wires and either negatively impact patient healing or lead to unnecessary procedures. Distortion is due to unequal magnification of different parts of an object. The purpose of this study was to assess distortion of three circular external fixator apparatuses with the transfixing wires at 30, 60, and 908. Distortion was greatest with all three apparatuses at 108 of rotation from parallel to the central X-ray beam. When comparing distortion between the three apparatuses, distortion was greatest with the transfixing wires at 308. The study authors concluded that distortion artifact is most severe when the transfixing wires are at smaller angles and when they are farther from the X-ray table. The circular external fixator should be placed in the center of, and parallel to, the primary X-ray beam and as close to the table as possible to reduce/prevent distortion artifact and possible radiographic misinterpretation of transfixing wire location.
ACCESSION #
111342936

 

Related Articles

  • Fractures in Single-humped Camels: A retrospective Study of 220 Cases (2008-2009). Ahmed Fathi Mohamed // Journal of King Abdulaziz University: Meteorology, Environment &;2012, Vol. 23 Issue 1, p3 

    The objective of this paper was to study the etiology and classification of fractures in dromedary camels and to evaluate their response to different treatment approaches. Two hundred and twenty cases of fractures in single-humped camels were admitted during the period from August 2008 to...

  • Treatment of Orthopaedic Problems with Manuflex® Disposable External Fixator in 15 Dogs and 7 Cats. ŞENEL, Oytun Okan; ERGİN, İrem; ÖZDEMİR, Özge; ULUSAN, Sinan; CSEBİ, Peter; DİOSZEGİ, Zoltan; BİLGİLİ, Hasan // Kafkas Universitesi Veteriner Fakultesi Dergisi;Sep/Oct2014, Vol. 20 Issue 5, p799 

    The aim of this study was to use Manuflex® disposible external fixator (MDEF), designed by two Hungarian orthopaedists, one of which is MD and the other is DVM, for treatment of orthopaedic problems, and to present the outcomes in dogs and cats. Cases included fractures of humerus (n=1),...

  • USO DE FIJADORES ESQUELÉTICOS EXTERNOS EN LA TÉCNICA DE DOBLE OSTEOTOMÍA PÉLVICA EN PERROS. Panta, José V.; Chipayo, Ysaac G.; Díaz, Diego C.; Fernández, Víctor A. // Revista de Investigaciones Veterinarias del Peru;dic2013, Vol. 24 Issue 4, p459 

    The aim of this study was to evaluate the use of external skeletal fixators as a means of stabilization, using the technique of double pelvic osteotomy in patients with hip dysplasia. Five crossbred dogs (3 males and 2 females), 3-4 months old, weighing 9.4 to 12.5 kg, and clinically healthy...

  • Use of external fixators for damage-control orthopaedics in natural disasters like the 2005 Pakistan earthquake. Awais, Syed; Saeed, Ayesha; Ch, Asad // International Orthopaedics;Aug2014, Vol. 38 Issue 8, p1563 

    Purpose: In the 2005 Pakistan earthquake, the great many injured with multiple fractures and open wounds provided a unique opportunity to practice damage-control orthopaedics. External fixators remain a time-tested tools for operating surgeons on such occasions. The locally manufactured, readily...

  • Temporary and definitive external fixation of war injuries: use of a French dedicated fixator. Mathieu, Laurent; Ouattara, Naklan; Poichotte, Antoine; Saint-Macari, Erwan; Barbier, Olivier; Rongiéras, Fréderic; Rigal, Sylvain // International Orthopaedics;Aug2014, Vol. 38 Issue 8, p1569 

    Purpose: External fixation is the recommended stabilization method for both open and closed fractures of long bones in forward surgical hospitals. Specific combat surgical tactics are best performed using dedicated external fixators. The Percy Fx© (Biomet) fixator was developed for this...

  • COMPLEX FRACTURES OF TIBIAL PLATEAU: ANYTHING NEW? MORRA, M.; D'AMBRA, R.; GUERRIERO, F. // Journal of Orthopedics;Jan-Apr2013, Vol. 5 Issue 1, p35 

    The purpose of this article is to highlight some controversial aspects in the diagnosis and treatment of these severe skeletal lesions. The use of a spanning external fixator should be adopted by all emergency departments. It is a method which allows immediate fracture stabilization and better...

  • Gradual bilateral genu varum correction in skeletal dysplasia using the Ilizarov method. Kim, Seung-Ju; Cielo, Balace; Song, Sang-Heon; Song, Hae-Ryong; Song, Sang-Yoon // Journal of Orthopaedic Science;Jul2011, Vol. 16 Issue 4, p405 

    Background: Options for lower limb realignment in skeletal dysplasia are acute versus gradual correction, internal versus external fixation, and external fixation with or without intramedullary nailing. The safety and versatility of the Ilizarov method in skeletal dysplasia patients makes it a...

  • Prevention of pin tract infection with iodine-supported titanium pins. Shirai, Toshiharu; Watanabe, Koji; Matsubara, Hidenori; Nomura, Issei; Fujiwara, Hiroyoshi; Arai, Yuji; Ikoma, Kazuya; Terauchi, Ryu; Kubo, Toshikazu; Tsuchiya, Hiroyuki // Journal of Orthopaedic Science;Jul2014, Vol. 19 Issue 4, p598 

    Background: Pin tract infection is one of the most common complications of external fixation. We developed techniques to coat titanium implant surfaces with iodine. This study clinically evaluated the infection-preventive effects and biological safety of iodine-coated external fixation pins....

  • Femoral Lengthening with Lengthening over a Nail has Fewer Complications than Intramedullary Skeletal Kinetic Distraction. Mahboubian, Shahab; Seah, Matthew; Fragomen, Austin; Rozbruch, S. // Clinical Orthopaedics & Related Research;Apr2012, Vol. 470 Issue 4, p1221 

    Background: Lengthening over a nail and internal lengthening nails have been developed to minimize or eliminate patients' time wearing a frame during femur lengthening. However it is unclear whether either of these two approaches results in faster times to union or fewer complications over the...

Share

Read the Article

Courtesy of THE LIBRARY OF VIRGINIA